92
action will cause the heaters to be turned off and the system will begin to cool.
Additionally, the turbo pump will be slowly turned off. In the screen below the
shutdown command has been activated. Notice that the turbo pump speed is
being reduced as the temperature is being cooled.
Should you wish to restart the system after you have activated the shutdown,
simply hit the reset button on the left side of the screen above. This will restart
the pumps and turn on the heaters.
After the temperature zones have cooled, turn OFF main power by placing switch
at rear of system into OFF (down) position. Manually vent the system for at least
5 minutes using the lever on front panel.
Diffusion Pump Shutdown
NOTE: During the SHUTDOWN procedure, the MS temperature zones are
reduced (<80°C). The system is manually vented to atmosphere when the trap
temperature is less than 80°C.
From System Control click Shutdown. The Shutdown display appears, click
Shutdown above the reset button.
The shutdown procedure and capabilities for the Diffusion Pump system are
similar to the turbomolecular system.
Shutdown Page
NOTE: The Diffusion Pump System shutdown sequence takes at least 30
minutes:
The Saturn GC/MS heaters are turned OFF.
The GC temperature zones should be cooled down before maintenance is
performed.
The diffusion pump heater is turned OFF while the Peltier Baffle cooler is
kept on. The DIFFUSION PUMP STATUS changes from NORMAL to
COOLING. After 15 minutes, when there is no pump vapor to contaminate
the system, the Peltier Baffle cooler is turned OFF. After another 15 minutes,