53
Item Description
Electron Multiplier Voltage Manual adjustment of the Final Gain Setting determine
from the Auto Tune check.
SIS Amplitude Adjust
Factor
Adjust each trap system for optimum ion sensitivity and
ejection. The voltage adjustment range is between 50 to
200%. The default is 100%.
Temperatures
Item Description
Start Bakeout Execute the bakeout conditions, which removes volatile
material from the ion trap surface.
Reset End a bakeout before the specified time.
Hide Keypad Hide the upper portion of the screen and expand the lower.
Analysis Conditions Set temperature for trap, manifold and transfer line. Click
Apply to initiate conditions.
Bake out Conditions Hold Time specifies the number of hours for bake out. Under
Trap, set the bake out temperature, such as 220 C. The
manifold bakeout temperature is fixed. The transfer line
bakeout temperature matches the analysis temperature.
Diagnostics Mode