TCD Detector
Item Description
Detector Oven On or off.
Turns the TCD oven on or off.
Electronics On or Off.
Indicates whether the installed TDC will be used or not.
Temperature (C) 50- 450 C.
Specifies the TCD Temperature.
Filament Temp (C) 0-390 C.
Specifies the Filament Temperature.
Time Initial - 999.99 min.
Range 0.5 and 5.0 mV.
Whenever possible, operate the TCD at the lowest practical
filament current.
Autozero Yes or No.
Set to yes at initial time means that the FID Autozero is on
continuous before the run starts. Autozero is automatically
disabled once the run starts.
Polarity Positive or Negative.
Converts the polarity to reverse the peaks that appear in the
negative direction.
Time Initial - 999.99 min.
Add Adds a line to the spreadsheet.
Insert Inserts a line above the currently selected row in the
spreadsheet.
Delete Deletes the currently selected row(s) in the spreadsheet.
Sort Sorts the spreadsheet rows by time.