To upload the Method from the 3900 GC to System Control, select “Upload Active
Method from Modules…” from the File menu.
You are then prompted for the name of the file under which to save the Method.
The active Method in System Control is updated to reflect the parameters
obtained from the 3900 GC.
The Startup Method
When you start System Control and display an Instrument Window, the last
active Method for that instrument is activated. When System Control is started, it
will return to the initial settings in the Method that was last used on the
instrument.
To change Methods, simply click on the Activate a Method button on the System
Control Toolbar or choose Activate Method… from the File menu.
Editing Methods from the 3900 Status and Control Window
The 3900 GC Status and Control Window contains a Method… button that allows
you to directly edit the active Method corresponding to that GC.