Customize Parameters window is separated into three sections. Ionization,
Isolation and Dissociation.
Ionization
Ionization Storage Level: The m/z value at which the rf is maintained during
ionization and the coarse isolation step. The range is 35 to 160 m/z, with a
default value of 48 m/z.
Ejection Amplitude: the amplitude (in volts) of the ejection waveform during the
coarse isolation step. The default value will work for most ions. For certain ions
that are less stable, this value may need to be reduced in order to minimize the
amount of energy imparted to the parent ion during isolation. The range of the
ejection amplitude is from 0 to 60 volts, with a default value of 20 volts.
Isolation
Low Edge Offset: The integer value step to optimize the ejection of the mass
just below the parent ion mass. The low edge offset range is 0 to 20 steps. The
default value is 6 .
Low Edge Offset affects the isolation window on the low mass side of the parent
ion. Increasing the value in the positive direction (increasing the default from 6 to
7 steps) opens the isolation window on the low mass side of the parent ion.
Decreasing the offset (decreasing from the default 6 to 5 steps) decreases the
window on the low mass side. The offset should be adjusted to minimize the
amplitude of the adjacent masses below the parent ion. Initially, adjust in 2 step
increments. Typically, a 2 step adjustment will eliminate the mass below the
parent ion.