Item Description
Delete Deletes the highlighted Time Events Table entry from the
table.
Sort Sorts the table in order of increasing start times.
Time Events
Event Description
Peak Width (WI) Changes the data bunching rate to allow for changing
peak widths during a run. If no WI events are
programmed, Data Handling will automatically adjust the
bunch rate as the peak widths increase or decrease. Any
time programming of peak widths prevents automatic
updating. Used to help reject noise spikes or to prevent
gradual baseline changes from being detected as peaks.
If a peak is being processed at the programmed time, the
event will not occur until a baseline segment is detected.
This bunching affects only how Data Handling views the
data; it does not have any effect on the raw data
bunching set in the Detector Information window for the
corresponding detector.
Inhibit Integrate (II) Suppresses integration of peak area or height. While II is
active, peak processing is disabled and no retention
times are displayed on the chromatogram or in the
results. In addition, any peak being processed at the start
of the II event will be aborted. Used to prevent integration
over a section of the chromatogram, sometimes to force
a baseline to be drawn. II disables peak sensing. In
contrast, SR is a post-run filter that rejects detected
peaks whose apices fall within the Solvent Reject
window.
Group Peaks (GR) Sums the areas of all peaks whose retention times fall
within the Group window. This causes all the peaks in
this window to be reported as a single peak with the
separation code GR and a retention time set to the
midpoint of the window. It is a post-integration function.
Used to treat all the peaks in a group as if they were a
single peak. The Peak Reject filter is applied after the
Group function, so that no peaks will be rejected from the
Group window because of size. A group peak is treated
like any other identified peak.
Solvent Reject (SR) Rejects all detected peaks whose apices fall within the
SR window. It is a post-integration filter. Used to exclude
peaks in a certain part of a chromatogram from the
report. Unlike the II event, SR can reject a mother peak
while retaining its tangent peak in the report.
Valley Baseline (VB) Forces peak baselines to be drawn through any valley
point that occurs during the VB window. It is a post-
integration function.