Viewing Calibration Curves
If you are performing calibrated calculations (external standard, internal standard
or normalized percent), you may view the calibration curves associated with the
active Method. Pull down the Results menu.
Select View Calibration Curves.
The Calibration Curve window is displayed if calibration data exists in the active
Method. If you have not yet performed any calibration runs with the active
Method, or if you have cleared calibration coefficients without adding any new
calibration data, a message box will tell you that no calibration data exists in the
Method.
You may print or export the selected curve or curves for all peaks in the peak table.
You may overlay curves with different fit parameters to determine the best type of fit
to use for your data.
You may manually enter calibration coefficients.
Right-click on a calibration point to exclude it from the curve fit calculation.