81
Control and Status
The Auto Tune methods are started from this display by clicking on the Start
Auto Tune button. The method maybe stopped by clicking on the Reset button.
The State
displays whether the Saturn MS is currently executing an auto tune
function or is idle.
The Function
shows which tuning function is currently being performed.
Method and Set Points
There are four auto tune checks to automatically set up, tune and calibrate the
mass spectrometer. Air/Water check, Electron Multiplier Tune, Mass Calibration
and Trap Function Calibration.
Set Points are for manual adjustment of the Electron Multiplier Voltage and
the SIS Amplitude Adjust Factor.
NOTE: The Air/Water Check will use the last 10
5
Electron Multiplier Setting and
not the manual setting. If the electron multiplier is replaced, the autotune of
Electron Multiplier Tune must be done before the Air/Water Check.
Operating Conditions
The operating conditions of the auto tune display will vary depending on the
currently executing auto tune function. The display will show the current results
of the function.