Item Description
Calibrate Vial 0 Position, Use
Mouse Buttons to Rotate
Carrousel
Clicking this button will display the following
dialog box.
Calibrate Vial 0 Position, Use
Mouse Buttons to Raise/Lower
Syringe
Clicking this button will display the following
dialog box.
Calibrate Injector Position, Start
Calibration
Press the Start Calibration. The CP-8400/8410
will rotate the tower so that it is over the
Injector position and lower the syringe sled so
that it is over the injector. If the needle guide
descends so that it is either resting on the
inject switch or below the needle cone on the
inject switch, use the Syringe Up button and
move the sled up until the needle guide is just
above the injector cone.
Calibrate Injector Position, Save
Calibration
Press the Save Calibration softkey when the
Injector calibration is complete.
Calibrate Injector Position, Use
Mouse Buttons to Rotate Tower
Position the tower directly over the injector
Inject Switch Locator cone. Clicking this button
will display the following dialog box.