Item Description
Rate (C/min) 1-200 ºC /min
The Rate in the first row is always blank and cannot be
edited.
Hold (min) 0.01-999.99 min
Total (min) 0.01-999.99 min
Cannot be edited.
Add Adds a line to the spreadsheet.
Insert Inserts a line above the currently selected row in the
spreadsheet.
Delete Deletes the currently selected row(s) in the spreadsheet.
Spilt Ratio Dialog Box
This dialog is displayed from the 3800 GC Injector window when a 1079 injector
is configured with type 1 EFC.
Item Description
Time 0.00-999.99 min
Split State On/Off
If the split state is ON, then the sample is split according to the split
ratio specified. If the split state is OFF, then all the sample enters
the column.
Split Ratio Off, 1 to 10,000
Use a split ratio of 100 after injection to vent the injector. Use a
very low split ratio after flushing to conserve carrier gas.
Add Adds a line to the spreadsheet.
Insert Inserts a line above the currently selected row in the spreadsheet.
Delete Deletes the currently selected row(s) in the spreadsheet.
Sort Sorts the spreadsheet rows by time.