Item Description
Vial I, II, or III
Selects which solvent vial to use for the solvent plug.
Volume (uL) 0 to 10.0 microliters with 10 uL syringe
0 to 5.0 microliters with 5 uL syringe
Specifies the amount of solvent to be drawn into the
syringe before the sample or internal standard.
Drawup Speed (uL/sec) 0.1 to 50.0 microliters per second with 10 uL syringe
0.1 to 25.0 microliters per second with 5 uL syringe
Specifies the speed at which the solvent will be
drawn up into the syringe.
Pause Time (sec) 0 to 9.9 seconds
Specifies how long the syringe is to remain in the
solvent vial after drawing up the solvent.
Air Gap Yes or No
If Yes, 1 microliter of room air will be drawn into the
syringe to create an air gap before the solvent plug.
User Defined Settings