Item Description
Excitation Storage
Level
Rf storage level in m/z when the dissociation waveform
is applied following isolation. The excitation storage level
range depends on the parent mass, but the storage level
must lie more than 2 mass units below the lowest
product ion value.
Excitation Amplitude Range for nonresonant excitation is 0 to 100 volts. For
resonant excitation, the range is 0 to 60 volts. Default
values for nonresonant excitation is 20 volts and 0.2
volts for resonant excitation.
Item Description
Ionization Storage Level The m/z value the rf is maintained at during
ionization and the coarse isolation. Range 35 to
160 m/z, default = 48 m/z.
Ejection Amplitude Amplitude (v) of the ejection waveform during
the coarse isolation step. Range 0 to 60 volts,
default=20.
Low Edge Offset Integer value step to optimize the ejection of
the mass just below the parent ion mass.
Range is 0 to 20 steps. The default value is 6 .
High Edge Offset The integer value step to optimize the ejection
of the mass just above the parent ion mass.
Range is -20 to 20 steps.