Whenever you execute changes, the event log at the bottom of the screen will be
updated with the time and the event that has occurred. The screen below is an
example.
At the conclusion of the bakeout, the trap, and manifold temperatures will return
to the set point values. New setpoint values can be applied during a bakeout. A
bakeout must be halted, by clicking the reset button, before a new bakeout
duration or new bakeout temperatures can be set.
Diagnostics Mode
Click Diagnostics in the top center of the screen.
The upper portion of the screen is divided into 3 sections: Control and Status,
Diagnostic Method, and Monitor States.
An automated diagnostics can be initiated by clicking on the Run to Completion
Box and clicking the Start