Axis BTA 2100D Manuel de l'opérateur Page 443

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 597
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 442
441
Internal Standard Calibration
When you use the Internal Standard calibration type, you calibrate with regular
calibration standards but you also add a known amount of another analyte, the
internal standard, to each calibration standard and sample. When you make an
injection, any variation in the injection volume will be reflected by a detectable
variation in the amount of the internal standard. Fluctuations may also be caused
by the process used to prepare the samples. In this case, adding the internal
standards before sample preparation corrects for recoveries. Your Peak Table
may contain as many as eight internal standard peaks. Each non-standard peak
refers to the internal standard peak to use for calculations. Having multiple
internal standards allows you to match each peak to a standard with similar
chemical properties, which should have a similar recovery.
In generating the curve for an Internal Standard Calibration, you plot ratios: the
peak size divided by the standard peak size on the y axis and amount of
standard divided by the amount of internal standard on the x axis. Even if your
injection volume varies slightly, the ratio of the sample to the internal standard
will remain constant.
Normalized Percent Calibration
The Normalize Percent calibration type gives a percentage amount of each
analyte relative to the total amount of all analytes in the sample, adjusted for
varying detector response to the different analytes. It is analogous to a regular
percentage calculation, except that the percentages are based on actual
amounts of sample components rather than on peak size. It is also analogous to
the Internal Standard calibration type in that the amounts are determined using a
calibration curve based on ratios of sample to internal standard. An alternative
method of generating ratios of amounts is to use either Internal Standard or
External Standard calibration types and choose Normalize Results in the
Integration Parameters window in the Data Handling section of the Method.
Choosing a Calibration Type
Your choice of which calibration type to use depends on your regulatory
requirements and sample workup techniques. If you can very accurately
measure your standard amounts, using an external standard calibration type is
the easiest. The Internal Standard calibration type accurately accounts for
variances in injection volume. You would also choose Internal Standard
calibration if you have varying loss during sample preparation. In this case you
would add a known amount of your Internal Standard compound before sample
preparation, and sample loss during preparation will be accounted for.
Vue de la page 442
1 2 ... 438 439 440 441 442 443 444 445 446 447 448 ... 596 597

Commentaires sur ces manuels

Pas de commentaire