Axis BTA 2100D Manuel de l'opérateur Page 263

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 597
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 262
261
Item Description
Autozero Yes or No.
Set to yes at initial time means that the TSD Autozero is
on continuous before the run starts. Autozero is
automatically disabled once the run starts.
Bead Power On or Off.
Time Constant Slow or Fast.
Add Adds a line to the spreadsheet.
Insert Inserts a line above the currently selected row in the
spreadsheet.
Delete Deletes the currently selected row(s) in the
spreadsheet.
Sort Sorts the spreadsheet rows by time.
EFC Type None or Type 12 for the TSD detector.
Make up Flow (ml/min) 0-50 ml/min.
H2 Flow (ml/min) 0-6.0 ml/min.
Air Flow (ml/min) 0- 200 ml/min.
Methanizer Installed Yes or No.
Methanizer Oven On or Off.
Temperature (C) 450 C.
Vue de la page 262