594
9 Peak 5.580 1.0000 1.4435 44.3 5.586 -0.001 297265
--- ------------ ---------- ========== ----- ------- ======= ======== ------
Totals: 14.5762 -0.002 3356238
Status Codes:
R - Reference peak
V - Out of verification tolerance
M - Missing peak
C - Out of calibration range
S - Internal Standard peak
Total Unidentified Counts : 0 counts
Detected Peaks: 7 Rejected Peaks: 0 Identified Peaks: 9
Standard Peak Amount:
2-Octanone Amount = 12
Multiplier: 1 Divisor: 1
Baseline Offset: -20 microVolts
Noise (used): 1 microVolts - monitored before this run
Manual injection
******************************************************************************
*
Item Description
Level Set in the SampleList or RecalcList (in System Control) or in the
Reintegration List (in Interactive Graphics). This corresponds to
the concentration level used in the verification sample, which can
be the same as one of the calibration levels, or can be between
levels used in calibration. The value in the peak table at this level
is assumed to be the amount of the compound used in the
verification sample.
Tolerance Set in the Verification dialog box accessed from the Peak Table
Window in the Data Handling section of the Method. This is the
maximum acceptable deviation percentage allowed before
considering a peak to be out of tolerance.
Expected
Result
This is the amount obtained from the peak table corresponding to
the designated verification level.
Calculated
Result
This is the result calculated from the calibration curve using the
detector response from the verification run. If no drift has occurred
in the system, and the amount of the compound in the verification
sample is identical to the amount of the same compound in the
calibration sample at the same level, then this value will be the
same as the expected result.
Dev. % The percentage that the calculated result deviates from the
expected result. If this value is larger than the tolerance, then a
verification failure occurs.