52
Item Description
Start Auto Tune Begins the event sequence that will tune the instrument. Only
checked boxes will be executed.
Reset Stops the auto tune process.
Continue Starts the next selected function if the single step option has been
checked.
Hide Keypad Causes upper portion of the screen to be removed so that the lower
portion can be expanded in the display.
Display
Selection
Select the display.
Method
Item Description
Air/Water Check Measures the air and water content of the trap, to
determine if there is a leak.
Electron Multiplier Tune Autoset the electron multiplier voltage.
FC-43 Mass Calibration Calibrate the mass axis. Choose Straight Line or FC-43.
Trap Function Calibration Calculate the waveforms for CI, MS/MS and SIS.
Single Step Observe each step in the auto tune process. Use Continue
to proceed after each step.
Set Points