Axis BTA 2100D Manuel de l'opérateur Page 183

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 597
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 182
181
Item Description
Isolation Time The dwell time during which the rf field is held
constant in the high mass ejection step.
Range 1 - 10 msec.
Excitation Time Time for collision-induced dissociation (CID)
by ion excitation. Range 0 to 1000 msec. The
default is 20 msec
Modulation Range Mass range over which the rf storage field is
modulated during resonant CID. The range is
0 to 12 steps, default is 2 steps. Value of 0
gives a fixed rf storage field, i.e., no
modulation.
Modulation Rate The number of microseconds that will be spent
at each step. The range is 29 usec to 5600
usec, default is 3000 usec/step.
Number of Frequencies Enter an odd integer in the range of 1 to 121.
A frequency number of one generates a single
supplementary frequency applied to the end
caps. An integer of 3, generates three
frequencies: the resonance frequency and
frequencies 500 Hz above and below that
frequency.
CID Frequency Offset To offset the dissociation frequency. The CID
frequency can be adjusted between 3000
Hz.
Automated Methods Development (AMD)
Item Description
Parent Ion Mass (m/z) Integer or exact mass. Parent ion mass range is 50 to
650 m/z.
Isolation Window (m/z) Integral and fractional mass isolation windows are both
accepted. Range is 1.0 to 14.0 m/z and is mass
dependent. The default value is 3.0 m/z.
Waveform Type Nonresonant or resonant excitation.
Vue de la page 182
1 2 ... 178 179 180 181 182 183 184 185 186 187 188 ... 596 597

Commentaires sur ces manuels

Pas de commentaire