From the Setup Ethernet Ports, click Setup for IP Address Management [BOOTP
Server].
Click Setup.
When the Setup BOOTP Server dialog box is displayed, turn on the 3800 or
3900 GC that has been connected to the network.
NOTE: If more than one Workstation is being connected on the same network,
only one Workstation should manage the IP addresses for all GCs connected (all
others should have the Manage IP addresses from this Workstation checkbox
unchecked
). If more than one Workstation is managing IP addresses, naming
conflicts may arise.
As the GC is powered on, an entry is added to the table. For each entry, an IP
address is automatically assigned. For each GC, you provide a host name that
will be used to identify it in the Varian MS Workstation.